User guide

Nebuliser with vibrating membrane

  • This document is for informational purposes only. Users must refer to the instructions for use (IFU) for further information on the use of Aerogen GO.

A small, portable* nebuliser

  • Lightweight, compact and can be powered by batteries or a mains adapter (2)
  • Equipped with a vibration membrane — a high-performance nebuliser technology (3)

Designed for a wide range of applications

  • From hospital to home, or on the move* (2)

The Aerogen Go uses vibration membrane technology to generate an aerosol (mist) from liquid medication, which is inhaled through a mouthpiece or face mask. (2)

Intended use

The Aerogen Go® Vibrating Mesh Nebuliser is designed for the nebulisation of approved medications for the treatment of respiratory conditions.

  • Only use medicines that have been prescribed for you by your doctor.

The Aerogen Go can be used by patients aged 6 months and older who require inhalation treatment and are able to breathe independently.

  • Children and persons requiring assistance or supervision must be supervised by an adult at all times when using the device.

Important information

  • The Aerogen Go is intended for use by a single patient. It should not be shared with other users.
  • Keep the Aerogen Go out of the reach of children unless it is used by children under the constant supervision of adults.
  • Do not drop your Aerogen Go or knock it against any surface. This could cause damage and void the warranty.
  • The Aerogen Go should be cleaned after each use and disinfected at the end of each day. For more information, see the ‘Cleaning and Disinfection’ section in this user manual.
  • This document is for informational purposes only. Users must refer to the instructions for use (IFU) for further warnings and important information.

Package contents

  • Main unit

  • Medication container

  • Mask for adults

  • Mask for children

  • Mouthpiece

  • Storage bag

  • Power supply type C
    type G (optional)

  • Alkaline batteries
    (2xAA/LR6)

Masks for adults and children are intended for those who cannot use the mouthpiece.

Setting up the Aerogen GO

  • Clean and disinfect your Aerogen Go thoroughly and carefully before using it for the first time. (For more information, see the ‘Cleaning and Disinfection’ section in this user manual.)

Inserting alkaline batteries

  • Unscrew the outer battery cover.

  • Open the inner battery cover.

  • Insert two AA alkaline batteries.

  • Press down on the inner battery cover until it clicks into place.

  • Turn the outer battery cover to close it.

  • Do not use multiple brands of batteries at the same time. Do not use old and new batteries at the same time. Replace all batteries at the same time.
  • If your Aerogen Go is likely to be stored for an extended period of time, remove the batteries and detach the medication container from the main unit.

Connecting to the power supply unit

Open the cover of the power supply connection and connect the power supply to this connection.

Plug the power supply into a suitable power outlet.

  • The Aerogen Go does not have a battery charging function.
  • Before using the power supply, remove the batteries from the main unit.
  • Do not touch the power supply with wet or damp hands, as this may result in electric shock.

Using the Aerogen Go®

Filling the medication

  • Connect the medication container to the main unit by pressing both parts together until you hear a click.

  • Turn the cap of the medication container counterclockwise to open it.

  • Fill the medication container with the recommended amount of medication according to your prescription.

  • Turn the cap of the medication container clockwise until it can no longer be turned to close it.

  • Before connecting, check that the silicone ring is correctly positioned on the main unit. After connecting, check that there is no gap between the two parts.
  • Do not touch the function membrane attached to the cap of the medication container (shown in light blue in the image above).
  • The minimum filling volume is 1 ml. Do not fill more than 10 ml, as indicated by the upper line on the side of the medication container.
  • Do not fill the medication container when it is not in use.

Inhalation therapy session

  • Connect either the mouthpiece or the mask to the medication container.

  • Pressing the on/off button starts the nebulisation process. When the green light is on, nebulisation has begun.

  • Sit upright and hold the Aerogen GO vertically. Breathe normally.

End of the inhalation session

When the medication is running low, the amount of aerosol decreases. Press the on/off button to switch off the device.
For safety reasons, the device switches off automatically after 10 minutes of nebulisation.

  • Do not cover the ventilation opening on the mouthpiece or mask.
  • Do not touch the inside of the aerosol opening, as this may damage the sensitive mesh mechanism inside.
  • Keep the main unit and power supply dry. If they get wet, stop using them immediately and wipe off the liquid.
  • If the device is not switched off at the end of the session, the nebuliser system may be damaged.

Cleaning and disinfection

Cleaning after each use

  • Remove the mouthpiece or mask and disconnect the medication container from the main unit by pressing the blue release button for the medication container on the back.

  • Dispose of any remaining liquid in the medication container.

  • Place the medication container, mouthpiece or mask in warm, filtered water with mild liquid washing-up liquid and shake a few times.

  • Rinse thoroughly with filtered water (from a bottle or tap). If stains remain after cleaning the parts, repeat steps 3 and 4.

  • Carefully shake off any excess water. Then leave to dry completely in the air on a clean, lint-free cloth. Once dry, place in the storage bag or another dry and clean container.

  • The main unit should not be washed. It can be wiped with a clean paper towel or cloth.
  • Ensure that the main unit or power supply does not come into contact with water or other liquids.
  • Do not use an oven, microwave or dishwasher to disinfect your Aerogen GO.
  • When handling the medication container, avoid touching the mesh inside the aerosol connection and the functional membrane on the cap.

Daily disinfection

Medication container, mouthpiece and mask for children/adults

After cleaning your Aerogen Go at the end of the day as described above, follow one of the recommended disinfection instructions below:

  • Using a commercially available baby bottle steriliser: Follow the instructions provided by the manufacturer of the commercially available steriliser. Do not use an oven, microwave or dishwasher to sterilise your Aerogen Go (1).

  • Use of alcohol:
    Immerse in 70–75% ethyl alcohol for approximately one minute (2). Carefully shake off excess liquid. Allow to air dry on a clean, lint-free cloth. Store in the storage bag or another dry and clean container.

  • Disinfecting the main unit
    Wipe the main unit with a clean cloth moistened with 70–75% ethyl alcohol.

Troubleshooting and support

Troubleshooting

If the Aerogen Go does not switch on after connecting the medication container to the main unit, check that the medication container has been connected to the main unit correctly. Viewed from the side (1), there should be no gap between the two parts. If necessary, reconnect the two parts and check that the silicone ring on the main unit (2) is seated correctly.

If you continue to experience problems with your Aerogen Go, please refer to the ‘Troubleshooting’ section in the instructions for use (IFU).

Contact your home care provider:

  • if the Aerogen Go still does not work after reading the instructions for use
  • to request a digital replacement copy of the instructions for use
  • to order replacement parts for the device and accessories
  • to make use of the warranty service.

For troubleshooting and warranty service, please provide the batch number of the medication container or the unique device identification (UDI) of the nebuliser's main unit.

Bei Fragen wenden Sie sich an Ihren Homecare Provider

* Lightweight (approx. 75 g without batteries); compact (74 x 46 x 96 mm); can be powered by batteries or a mains adapter.

1. Aerogen-archived data. 2. EU-30-2173 Rev. 01 Aerogen Go - user manual
3. Pritchard JN, et al. Further delivery. 2018;9(2):121-136